Serinus® 10 User Manual 3.2
The calibration will automatically run again after the specified
amount of time. This field specifies the delay period (from
1 - 20,000 units, as specified below).
This is where the user can define the type of units for the Repeat
delay period. For example, A Repeat of “3” and units of “Days”
means that a calibration will automatically be performed every
three days.
When Enabled the instrument will automatically perform a Span
Calibrate O3 at the end of the Cycle Time and adjust the gain
based on the Span Level.
When Disabled it will do a precision check only, no adjustment is
made.
Timed calibration with span compensation enabled does not fulfil
U.S. EPA approval or EN certification.
Enter the concentration of span gas expected. Used when the Span
Compensation is Enabled.
3.4.11 Pressure Calibration Menu
Figure 40 – Pressure Calibration Menu Screen
Main Menu → Calibration Menu → Pressure Calibration Menu
Entering this menu will set the valves to the pressure calibration configuration; leaving the menu will
restore the valves to normal operation (refer to Section 5.3).
The zero point for the calibration. Activating this item will open a dialog
box of instructions.
The high point for the calibration. Activating this item will open a dialog
box of instructions.
Select the units that the pressure will be displayed in (torr, PSI, mBar,
ATM or kPa).
The current ambient pressure.
The current pressure in the optical cell.
The current optical cell pressure displayed as a raw voltage.