Serinus® 10 User Manual 3.2
3.4.4 Status Menu
Main Menu → Analyser State Menu → Status Menu
The Status Menu presents a list of the current Pass/Fail status of the main components. During
warm-up, the status of some parameters will be a dashed line.
Figure 33 – Status Menu Screen
This field enters a screen with a log of all the events that the instrument
has performed. These events include errors and warnings. This log is
stored on the USB memory stick. The log is organised by month. When
you enter this screen you will be prompted to enter the month for
which you wish to view events.
This field allows the user to display the list of current errors and
warnings on the screen.
This field is visible with the next service due date if the service is due
within the next two weeks.
Error if the system is “In Maintenance” (refer to Section 3.4.14).
Pass if the +5 V power supply is within the acceptable range.
Pass if the +12 V power supply is within the acceptable range.
Pass if the analog power supply is within the acceptable range (+12 V).
Pass if the analog power supply is within the acceptable range (-12 V).
Fail only if a problem is detected with the analog to digital conversion.
Ozone Lamp Temp.
[IZS option]
Pass if the Ozone generator lamp is the correct temperature.
Pass if the UV lamp heater temperature is within ± 10% of the heater
set point (refer to Section 3.4.5).
Check if the UV lamp current is within acceptable limits 8 - 12 mA.
Checks that the reference voltage is within acceptable limits 1.5 - 4.5 V.
Pass if the system has an adequate electrical supply.
Error if the electronics are in Diagnostic Mode (refer to Section 3.4.15).
Error if the Pres/Temp/Flow Comp. is disabled (refer to Section 3.4.15).