Indicates the gas flow through the sample port of the instrument.
The value should be around 0.50 slpm. If there is an error with the
sample flow, it will read 0.00 slpm.
3.4.7 Voltage Menu
Figure 36 – Voltage Menu Screen
Main Menu → Analyser State Menu → Voltage Menu
Voltage from the sensor proportional to the detected signal from
the optical cell. This voltage represents the actual measurement of
gas.
Displays the detector voltage after PGA scaling.
A voltage offset from the detector that is removed to measure the
concentration voltage. This is set to 3 V (using the input pot) at
start-up and automatically adjusts if the reference voltage drops
below 2 V or above 4 V.
Flow Voltage
[Internal Pump Option]
The current voltage measured from the sample flow.
O3 Gen. Current
[IZS Option]
+12 V (primary) power supply. The value should be within ± 2 V.
-12 V (primary) power supply. The value should be within ± 2 V.