Operation
Undiluted O3
Displays the actual reading on the analyser before the dilution
ratio is applied. If the Dilution ratio is 1 both undiluted O3 and
diluted O3 field will be the same value.
Dilution Ratio
Entering a value here will multiply the displayed and recorded
measurements by the dilution amount. For example, if the
instrument is measuring a source where the average concentration
is above the upper limit of the measurement range, a dilution
probe with a fixed ratio dilution ratio can be used to reduce the
level measured by the instrument, so for a 4:1 dilution ratio, enter
a value of 4.
Enter the ratio here so the instrument can display the correct
value on the instrument.
The default value is 1.00 (this indicates no dilution is applied).
Diluted O3
Displays the current reading on the analyser after the dilution ratio
is applied. If the Dilution ratio is 1 both undiluted O3 and diluted
O3 field will be the same value.
3.4.23 Communications Menu
Figure 52 Communications Menu Screen
Main Menu Communications Menu
Configures how the instrument communicates with external instrumentation and data loggers.
Data Logging Menu
Refer to Section 3.4.24.
Serial Communication Menu
Refer to Section 3.4.25.
Analog Input Menu
Refer to Section 3.4.26.
Analog Output Menu
Refer to Section 3.4.27.
Digital Input Menu
Refer to Section 3.4.28.
Digital Output Menu
Refer to Section 3.4.29.
Network Menu
[optional]
Refer to Section 3.4.30.
Bluetooth Menu
Refer to Section 3.4.31.