Serinus® 10 User Manual 3.2
5.3.3 Pressure Calibration (Internal Pump Option Only)
The internal pump requires a separate pressure calibration procedure that replaces the one used in
Section 5.3.1 and 5.3.2.
Note: Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
Note: Ensure units of measure are the same on both the barometer and instrument.
Equipment Required
• Barometer
Procedure
1. Disconnect any external tubing connected to the rear of the instrument.
2. Open - Main Menu → Calibration Menu → Pressure Calibration - (read displayed note) - OK.
Note: Entering this menu will switch OFF the internal pump and place the valve sequencing on
hold; normal sampling will be interrupted.
3. Connect a barometer to the Sample Port (refer to Figure 105).
Figure 105 – Pressure Calibration (Internal Pump Option Only) Set-up
4. Edit - Vacuum Set Pt. - (Read displayed instructions) - OK.