5.3.3 Pressure Calibration (Internal Pump Option Only)
The internal pump requires a separate pressure calibration procedure that replaces the one used in
Section 5.3.1 and 5.3.2.
Note: Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
Note: Ensure units of measure are the same on both the barometer and instrument.
Equipment Required
Barometer
Procedure
1. Disconnect any external tubing connected to the rear of the instrument.
2. Open - Main Menu Calibration Menu Pressure Calibration - (read displayed note) - OK.
Note: Entering this menu will switch OFF the internal pump and place the valve sequencing on
hold; normal sampling will be interrupted.
3. Connect a barometer to the Sample Port (refer to Figure 105).
Figure 105 Pressure Calibration (Internal Pump Option Only) Set-up
4. Edit - Vacuum Set Pt. - (Read displayed instructions) - OK.