Instrument Gain
This is a multiplication factor which is used to adjust the
concentration measurement to the appropriate level (set by
performing a Span Calibrate O3). This should be recorded after each
calibration in the station log book.
Next Service Due
A field that notifies the user when the next instrument service is due.
This value is editable in the Next Service Due field of the Advanced
Menu (refer to Section 3.4.34). This field is only displayed in the two
weeks prior to the date displayed in this field or after the date has
occurred.
3.4.2 Main Menu
There are six menus on the Main Menu screen.
Figure 31 Main Menu screen
Analyser State Menu
Refer to Section 3.4.3.
General Settings Menu
Refer to Section 3.4.8.
Measurement Settings Menu
Refer to Section 3.4.9.
Calibration Menu
Refer to Section 3.4.10.
Service Menu
Refer to Section 3.4.14.
Communications Menu
Refer to Section 3.4.23.
Trend Display Menu
Refer to Section 3.4.32.