2.4 U.S. EPA Equivalent Set-Up
The Serinus 10 is designated as equivalent method EQOA-0809-187 by the U.S. EPA (40 CFR Part 53).
The Serinus 10 must be used under the following conditions to satisfy its equivalency:
Range
0 - 500 ppb
Ambient Temperature
20 - 30 °C
Line Voltage
105 - 125 VAC, 60 Hz
Pump
Ecotech optional internal or external pump
Filter
Factory setup to meet requirement:
Instrument Settings
If the units in the measurement menu are changed from volumetric to gravimetric or gravimetric to
volumetric, the instrument must be re-calibrated.
The following menu selections must be used:
Calibration Menu
Span Comp: Disabled
Diagnostics Menu
Pres/Temp/Flow Comp: ON
Diagnostic Mode: Operate
Control Loop: Enabled
The instrument must be operated and maintained in accordance with this user manual.
The Serinus 10 Analyser is designated U.S. EPA equivalent method with or without the following
options/items:
Internal pump
Rack mount assembly
Internal zero/span assembly (IZS)
Optional Ethernet port