Serinus® 10 User Manual 3.2
Page 218
Table 29 Failure Bit Map (Positive Logic)
Failure Bit
Meaning if set to 1
0
Flow sensor failure.
1
Instrument failure. Note that while the In Maintenance mode reports as an instrument
failure with a red light on the front panel, for Bayern-Hessen this particular error is
merely a status instead of a failure.
2
-
3
Lamp failure (S40 family only).
4
-
5
Cell heater failure (S30, S40 and S50 family only).
6
-
7
-
ST
Set the instrument mode.
Command Format
<STX>{ST}{< kkk>}<SP>{command}<ETC><bcc1><bcc2>
Where:
kkk Device’s Serial ID. This field is optional, but if provided it must be padded
with zeros to be 3 characters long. The value must match one of the
following: the instrument’s Bayern-Hessen ID, 000, or ??? (three
question marks).
Command M, N or K for Measure, Zero or Span mode.
bcc1 First byte of the block check calculation.
bcc2 Second byte of the block check calculation.
Device response
The device does not issue a response to this command.