bcc2 Second byte of the block check calculation.
Device response
The instrument responds with a variable length string, depending on how many measured gasses have
been assigned an ID above 0. The text between the [ ] will be repeated once for each reported gas.
<STX>{MD}{cc}[<SP><kkk><SP><+nnnn+ee><SP><ss><SP><ff><SP><mmm><SP>eeeeee<SP>]<ETC><
bcc1><bcc2>
Where:
<SP> Space (0x20 hex).
cc The number of gasses reported (0..5). The text in between the [ ] will be
repeated once for each gas reported.
kkk The Bayern-Hessen instrument ID.
+nnnn+ee Gas concentration.
ss Status byte (see table below for individual bits).
ff Failure byte (see table below for individual bits).
mmm Gas ID.
eeeeee Ecotech instrument ID.
bcc1 First byte of the block check calculation.
bcc2 Second byte of the block check calculation.
Table 28 – Status Bit Map