Operation
Diagnostic Control
Error if the control loop is disabled (refer to Section 3.4.15).
Valve Manual Control
Error if the valves have been placed in manual control mode (refer to
Section 3.4.18).
O3 Gen. Manual Control
[IZS option]
Error if the Ozone generator is under manual control (refer to
Section 3.4.12).
O3 Conc V Saturated
Indicates if the voltage of the concentration during measurement is
within the limits of the analog to digital converter (-0.26 to 3.29 V).
BkGnd Conc V Saturated
Indicates if the voltage of the concentration during background
measurement is within the limits of the analog to digital converter
(-0.26 to 3.29 V).
Ozone Gen. Cal
[IZS option]
Fail if the instrument is performing an O3 generator calibration (refer to
Section 3.4.12).
Pressure Calibration
Error if the user is performing a pressure calibration.
Flow Calibration
Error if the user is performing a flow calibration.
Flow Fault
Ok when the instrument has acceptable sample flow based on the
difference between cell and ambient pressures.
With the internal pump option this fault is monitored by a flow sensor.
Cell Press > Ambient
This error occurs when the pressure inside the instrument exceeds the
ambient pressure.
Flow Block Temp.
[Internal pump option]
Pass if the flow block temperature is within 10% of the heater set point
(to keep a constant accurate flow).
Chassis Temp.
Pass if the chassis temperature is within the acceptable limits
(0 - 50 °C).
USB Stick Disconnected
Detects whether a USB memory stick is plugged into the front USB port.
Instrument Warmup
Ok once the instrument is out of warm-up status.
3.4.5 Temperature Menu
Figure 34 Temperature Menu Screen