6. Now that we have isolated the lamp block, remove the screw to dismantle the lamp mounting
block.
Note: Be careful while dismantling lamp mounting block Quartz Window may fall on the ground.
Figure 140 Remove Lamp Mounting Block
7. Remove the Kynar elbow fitting from the lamp block by turning counter-clockwise.
Figure 141 Remove Kynar Elbow Fitting
8. Remove the spring.
9. Remove the orifice by screwing in the orifice removal tool (finer thread end) clockwise, then gently
pulling out the orifice.