EasyManua.ls Logo

Ecotech Acoem Serinus 10 - Ozone Scrubber Check; Figure 128 - Barometer to Calibration Port

Ecotech Acoem Serinus 10
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus® 10 User Manual 3.2
Page 150
22. Connect the barometer to Calibration Port.
Figure 128 Barometer to Calibration Port
23. Allow the internal pump time to evacuate the pneumatic system (the time required will depend on
the condition of the pump, 1 - 2 minutes).
24. Open - Main Menu Service Menu Diagnostics Menu Internal Pump Menu.
25. Toggle Off - Internal Pump Off.
26. Note the value on the barometer. Wait for three minutes, the value should not drop more than
5 kpa (37.5 torr). If the leak check passed skip to step 29.
27. If you passed the first leak check on the sample side then the leak is between the calibration valve
manifold and the calibration port.
28. Repair and re-run the leak check procedure.
29. Slowly release the vacuum by gradually loosening the barometer on the Calibration port until the
instrument has returned to ambient pressure.
30. Remove the blocker on the pump bypass flow.
31. Open - Main Menu Service Menu Diagnostics Menu Valve Menu.
32. Enable - Valve Sequencing Enabled.
33. Open - Main Menu Service Menu Diagnostics Menu Internal Pump Menu.
34. Select - Pump Control Start - Accept.
6.4.4 Ozone Scrubber Check
The performance of the Ozone scrubber is critical to the Serinus 10. Although the Ozone scrubber will
theoretically last forever (if only exposed to clean air and Ozone), exposure to other elements in the
atmosphere will adversely affect the life span of the scrubber. A weak or failed scrubber can result in
noisy measurements, frequently caused by excessively high gain.

Table of Contents

Related product manuals