Operation
Vacuum Cal Mode
Defaults to Off. When turned On, the valves will be set to the same
state as during a Vacuum Set Pt. adjustment, but there is no
adjustment. Used for checking the accuracy of the vacuum pressure
calibration. There are no dialog boxes or prompts, so the user needs to
follow similar steps and precautions as during Vacuum Set Pt.
Ambient Cal Mode
Defaults to OFF. Similar to Vacuum Cal Mode, except the valves are set
to check the ambient calibration.
3.4.12 Ozone Calibration Menu (Option)
Figure 41 Ozone Calibration Menu Screen
Main Menu Calibration Menu Ozone Calibration Menu
This menu only appears when the IZS option is installed (refer to Section 8.6).
Ozone Set Point
This is the amount of Ozone that will be generated by the IZS
during an internal span calibration.
Ozone Calibration
Selecting Start on this field will begin an Ozone generator
calibration. This process will take at least 25 minutes and can take
up to 5 hours. If a calibration is running, this field will display which
step it is on (1 - 5) and pressing Stop will terminate the calibration.
An Ozone generator calibration can be interrupted at any time
(however, restarting the calibration will always begin at the first
step).
If the user has the internal pump option and changes the flow set
point, then the user will need to recalibrate the Ozone generator.