Serinus® 10 User Manual 3.2
3.4.35 Hardware Menu
Advanced Menu → Hardware Menu
Figure 64 – Hardware Menu Screen
This menu contains factory hardware installations. If the user reset to factory defaults then they may
need to revisit this menu to enable your installed optional features.
Select the instrument model. Normally this only needs to be
reset when the configuration is corrupted. The selections
available will depend on licensing. It is not recommended to
run an instrument with firmware set to an incorrect model.
Choosing the incorrect front panel will result in the traffic
lights behaving inconsistently. Default is Aluminium.
When Enabled indicates the instrument has a network port
installed. Default is Disabled.
Orifice Size
[Internal Pump Option Disabled]
Specify the input orifice for instruments without an internal
pump. Default is 0.5.
When Enabled indicates the instrument has an Internal pump
installed. Default is Disabled.
When Enabled indicates the instrument has the Internal Zero
module option installed. Default is Disabled.
When Enabled indicates the instrument has the Internal Zero
Span option installed. Default is Disabled.
When Enabled indicates the instrument has a shielded bench
installed. Default is Disabled.
CAUTION
No items in this menu should be accessed without authorization and supervision
of qualified service personnel.