Ecotech Acoem Serinus 10 - Figure 132 - Loosening Tube Collet from Detector End; Figure 133 - Removing Two Screws

Ecotech Acoem Serinus 10
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service
Procedure
Obey below procedure to clean the optical cell:
1. Turn OFF the Vacuum source and allow the instrument to return to ambient.
2. Turn OFF the instrument.
3. Open the lid of the instrument to access the Optical Bench (refer to Section 2.1, Opening the
Instrument).
4. Loosen the tube collet from the detector side of the measurement cell cover (Optical Cell).
Figure 132 Loosening Tube Collet from Detector End
5. Remove the two screws holding the detector block to the Optical Bench.
Figure 133 Removing Two Screws