Serinus® 10 User Manual 3.2
3.4.13 Flow Calibration Menu (Option)
Figure 42 – Flow Calibration Menu Screen
Main Menu → Calibration Menu → Flow Calibration Menu
This menu only appears when the internal pump option is installed. Refer to Section 5.8 for the
calibration procedure. Setting the Pump Control to MANUAL will automatically disable the valve
sequencing ready for flow calibration.
The desired sample flow. If the user has the IZS option and
changes the Flow Set Point, then the user will need to
recalibrate the Ozone generator.
The current pressure in the measurement cell.
The flow at which the last flow calibration was performed. If
Pump Control is set to MANUAL and Internal Pump is ON, this
field can be edited to calibrate the current flow.
Calibration must be done at or near the flow set point for best
results (refer to Section 5.8).
If Pump Control is set to MANUAL and Internal Pump is OFF, this
command will calibrate the flow sensor zero point (refer to
Section 5.8).
Current sample flow through the instrument. This is only
accurate when reading close to the flow calibration point.
This field allows the internal pump to be turned ON or OFF. This
field is only editable when the Pump Control field is set to
MANUAL.
Set to MANUAL to disable the automatic pump control. AUTO
allows the flow PID to modify the pump coarse and fine settings.
START will transition to AUTO after one second.
Internal pump speed control (Coarse). This field is only editable
when the Pump Control field is set to MANUAL.
Internal Pump speed control (Fine). This field is only editable
when the Pump Control field is set to MANUAL.