Serinus® 10 User Manual 3.2
Page 74
Background/Measure
Indicates if the Background/Measure valve on the Calibration
Valve Manifold is Off or On. This will determine the port the
instrument draws its sample from.
Off = Flow from NO to C (drawing the sample through the O3
Scrubber into the Optical Cell).
On = Flow from NC to C (drawing the sample directly into the
Optical Cell).
When Open, the sample feeds directly into the measurement cell;
when Closed, the sample feeds through the Ozone scrubber before
filling the measurement cell.
Pressurised Zero
[optional]
Indicates if the optional pressurised zero port valve is Off or On
(refer to Section 8.5).
Internal Span A
[IZS option]
Controls the internal/external span and zero gas (refer to
Section 8.6).
Internal Span B
[IZS option]
Controls the internal/external span and zero gas (refer to
Section 8.6).
3.4.19 Tests Menu
Figure 48 Tests Menu Screen
Main Menu Service Menu Diagnostics Menu Tests Menu
Screen Test
Performs a screen test by drawing lines and images on the screen
so that the operator can determine if there are any faults in the
screen. Press a keypad key to step through the test.
The up and down arrow keys will adjust the contrast.
Digital Input Test Menu
Refer to Section 3.4.20.
Digital Output Test Menu
Refer to Section 3.4.21