Service
If a faulty Ozone scrubber is suspected perform the following steps:
Equipment Required
Reference Ozone Scrubber
Span Source of 400 ppb Ozone
Procedure
1. Connect a span source (approximately 0.400 ppm O
3
) to the sample port, allow the instrument to
stabilise. Then record the response.
Figure 129 Connect Span Source
2. Replace the scrubber with a reference Ozone scrubber (refer to Section 6.4.4.1). Allow the
instrument to stabilize and record the response.
3. Compare the two readings. If the second reading exceeds the first by more than 10% the scrubber
should be replaced.
6.4.4.1 Scrubber Replacement
Figure 130 Ozone Scrubber
1. Turn OFF the Vacuum source and allow the instrument to return to ambient.