Serinus® 10 User Manual 3.2
Page 182
8.6.4 Serinus 10 Schematic with IZS Option
Figure 158 IZS Schematic
8.6.5 Ozone Generator Characterisation
This procedure allows the instrument to characterise the Ozone generation of the lamp to provide
accurate internal span points. This should be done whenever the lamp is adjusted, changed or when
the generated output of Ozone no longer matches the set point. If the user has the internal pump
option installed, they will need to run the Ozone calibration every time the target flow is changed.
Equipment Required
N/A
Procedure
1. Ensure the instrument has fully warmed up (minimum of one hour), that there are no status faults
and the instrument functioning normally.
2. Open - Main Menu Calibration Menu Ozone Calibration Menu.
3. Start - Ozone Calibration - (the instrument will now automatically step through five calibration
points. Step 1 - Step 5).
Note: Pressing Stop or terminating the O3 Gen/Photometer point will terminate the calibration
without saving any changes.
Note: The process will take at least 25 minutes and may take up to 5 hours. Ensure that the
instrument has been running for at least one hour before any calibration is performed to ensure
sufficient stability.