Serinus® 10 User Manual 3.2
8.5 High Pressurised Zero Valve (PN: E020109)
Note: This is not intended as a source for calibrating the instrument. This should only be used to
do a precision check (e.g. Level 2 calibration).
Refer to Section 5.9 for operation of this installed option.
8.6 Internal Zero & Span (IZS) (PN: E020130)
Note: This is not intended as a source for calibrating the instrument. This should only be used to
do a precision check (e.g. Level 2 calibration).
Note: Serinus 10 with both IZS and Internal Pump options is currently under development. Contact
Ecotech for more information.
The Serinus 10 can be fitted with an internal Ozone generator and Ozone scrubber to provide span gas
and zero air for calibration. This calibration can be automated and negates the need for an external
Ozone source. An additional port (Aux-In) is utilised to provide source air for the scrubber and Ozone
generator
8.6.1 Hardware Setup
This procedure will need to be followed after a factory reset.
Procedure
1. Press - (the green instrument status light button), this will take user to the home screen.
2. Press - (-99+) on the keypad. This will open the Advanced Menu.
3. Open - Advanced Menu → Hardware Menu.
4. Enable – IZS (Zero & Span) → Enabled.
8.6.2 IZS Additional Components
The Serinus 10 IZS includes the following components:
Table 14 – IZS Components
Fitting Kynar Bulkhead Union
Adaptor, 1/4" Fitting to 1/8"
Scrubber Assembly Charcoal
Manifold Assembly, IZS, Ozone
Ozone generator Assembly, S10
Scrubber Mounting Bracket