Connect the instrument’s power cord into the instrument and mains power outlet.
Figure 18 – Connect Power Cord
Turn ON the power switch.
Figure 19 – Turn ON Power Switch
2.3.2 Pneumatic Connections
The Serinus® 10 has three pneumatic ports on the back panel of the instrument, the Sample Port, the
Calibration Port and the Exhaust Port. All tubing and fittings used should follow the instructions below:
Must be made of Teflon® FEP material, Kynar®, stainless steel, glass or any other suitably inert
material.
Sample line should not be more than two meters in length with ⅛ inch ID and ¼ inch OD.
Sample inlet pressure should not exceed 5 kPa above ambient pressure.
Tubing must be cut accurately and any burrs removed.