Analog Inputs
The instrument is also equipped with three analog voltage inputs (0 - 5 VDC CAT 1) with resolution of
15 bits plus polarity.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Digital Status Inputs
The instrument is equipped with eight logic level inputs (0 - 5 VDC CAT 1) for the external control of
zero/span calibration sequences.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Digital Status Outputs
The instrument is equipped with eight open collector outputs, which will convey instrument status
conditions and warning alarms such as no flow, sample mode, etc.
CAUTION
Exceeding 12 VDC or drawing greater than 400 mA on a single output or a total
greater than 2 A across the eight outputs can permanently damage the
instrument and void the warranty.
Bluetooth
This allows for remote access of the instrument to any Android device with the Serinus Remote
Application installed on it. It uses Bluetooth to control the instrument, view parameters, download
data and construct real-time graphs.