6 Configure advanced options for process
measurement
6.1 Configure Response Time
Display Menu > Configuration > Process Measurement > Response Time
ProLink III Device Tools > Configuration > Process Measurement > Response Time
Field Communicator Not available
Response Time controls the speed of various internal processes that are involved in retrieving electronic data
from the sensor and converting it to process data.
Response Time affects all process and diagnostic variables.
Restriction
Response Time is configurable only if you are using the enhanced core processor. If you are using the
standard core processor, Response Time is set to Low Filtering and cannot be changed.
Procedure
Set Response Time as desired.
Option
Description
Normal Appropriate for typical applications.
High Filtering Slower response. Appropriate for applications with significant amount of entrained gas or
process noise.
Low Filtering Fastest response. Appropriate for proving or filling applications.
Service Do not select unless directed by Micro Motion personnel.
6.2 Detect and report two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues.
The transmitter provides two methods to detect and report or respond to two-phase flow.
6.2.1 Detect two-phase flow using density
Display
Menu > Configuration > Process Measurement > Density
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Slug Low Limit
Configure > Manual Setup > Measurements > Density > Slug High Limit
Configure > Manual Setup > Measurements > Density > Slug Duration
Configuration and Use Manual Configure advanced options for process measurement
MMI-20025166 March 2019
Configuration and Use Manual 105

Table of Contents

Other manuals for Emerson Micro Motion 5700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion 5700 and is the answer not in the manual?

Emerson Micro Motion 5700 Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion 5700
CategoryTransmitter
LanguageEnglish

Related product manuals