• Default: None
Options for Discrete Input Action
Action Label
Display ProLink III Field Communicator
Standard
Start sensor zero Start Zero Calibration Start Sensor Zero Start Sensor Zero
Totalizers
Start/stop all totalizers and
inventories
Start/stop all totalizers Start or Stop All Totalizers Start/Stop All Totals
Reset totalizer X Reset Total X Totalizer X Reset Total X
Reset all totalizers and
inventories
Reset All Totals Reset All Totals Reset All Totals
Concentration measurement
Increment CM matrix Increment Matrix Increment ED Curve Increment Curve
Batching
Begin batch Begin Batch Begin Batch Start Batch
End batch End Batch End Batch End Batch
Pause batch Pause Batch Pause Batch Pause Batch
Resume batch Resume Batch Resume Batch Resume Batch
Increment batch preset Increment Preset Increment Batch Preset Increment Preset
Inhibit batch totalizer Inhibit Totalizer Inhibit Batch Totalizing Inhibit Batch Totalizing
Inhibit batch Inhibit Batch Inhibit Batch Start Inhibit Batch Start
Inhibit batch flow Inhibit Flow Allow Batch End with Flow Allow Batch End with Flow
8.7.2 Configure Discrete Input Polarity
Display
Menu > Configuration > Inputs/Outputs > Channel x > I/O Settings > Polarity
ProLink III Device Tools > Configuration > I/O > Inputs > Discrete Input > Discrete Input x Polarity
Field Communicator Configure > Manual Setup > Inputs/Outputs > Channel x > Discrete Input x > DI x Polarity
A Discrete Input has two states: ON and OFF. Discrete Input Polarity controls how the transmitter maps the
input voltage to the ON and OFF states.
Procedure
Set Discrete Input Polarity as desired.
• Default: Active Low
Configuration and Use Manual Integrate the meter with the control system
MMI-20025166 March 2019
Configuration and Use Manual 163