17.9 Frequency Input problems
17.9.1 No Frequency Input
Cause
Wiring problem
Problem with external device
Channel not configured for desired output or input
Channel not licensed
Recommended actions
1. Verify that the external device is operating correctly.
2. Test the output from the external device.
3. Verify the wiring between the transmitter and the external device.
4. If applicable, ensure that the loop is powered.
5. Verify that the channel is wired and configured as a Frequency Input.
6. Purchase a license for the channel and upgrade the transmitter firmware.
17.9.2 Frequency Input reading consistently wrong
Cause
Incorrect configuration
Recommended actions
Check the scaling of the Frequency Input.
17.9.3 Frequency Input reading slightly inaccurate
Cause
External device not calibrated to match transmitter
Recommended actions
Calibrate the external device against the transmitter reading.
17.10 Frequency Output problems
17.10.1 No FO
Cause
Stopped totalizer
Process condition below cutoff