• Using ProLink III: Choose Device Tools > Configuration > I/O > Inputs > Action Assignment. Then
assign the desired action to the desired enhanced event.
• Using the Field Communicator: Choose Configure > Alert Setup > Enhanced Events. Then assign
the desired action to the desired enhanced event.
Options for Enhanced Event Action
Action Label
Display PLIII FC
Standard
Start sensor zero Start Zero Calibration Start Sensor Zero Start Sensor Zero
Totalizers
Start/stop all totalizers and
inventories
Start/stop all totalizers Start or Stop All Totalizers Start/Stop All Totals
Reset totalizer X Reset Total X Totalizer X Reset Total X
Reset all totalizers and
inventories
Reset All Totals Reset All Totals Reset All Totals
Concentration measurement
Increment CM matrix Increment Matrix Increment ED Curve Increment Curve
Batching
Begin batch Begin Batch Begin Batch Start Batch
End batch End Batch End Batch End Batch
Resume batch Resume Batch Resume Batch Resume Batch
Increment batch preset Increment Preset Increment Batch Preset Increment Preset
Inhibit batch totalizer Inhibit Totalizer Inhibit Batch Totalizing Inhibit Batch Totalizing
Inhibit batch Inhibit Batch Inhibit Batch Start Inhibit Batch Start
Inhibit batch flow Inhibit Flow Allow Batch End with Flow Allow Batch End with Flow
6.5 Configure totalizers and inventories
Display
Menu > Configuration > Process Measurement > Totalizers & Inventories
ProLink III Device Tools > Totalizer Control > Totalizers
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > Configure Totalizers
The transmitter provides seven configurable totalizers and seven configurable inventories. Each totalizer and
each inventory can be configured independently.
Totalizers track the process since the last totalizer reset. Inventories track the process since the last inventory
reset. Inventories are typically used to track the process across totalizer resets.
Configure advanced options for process measurement Configuration and Use Manual
March 2019 MMI-20025166
110 Micro Motion Model 5700 Transmitters with Configurable Outputs