Postrequisites
View the results and take any appropriate actions.
14.1.2 View SMV test results
After each SMV test, the pass/fail result is displayed automatically. Detailed results are also available for
licensed versions.
Tip
For licensed SMV versions only, when you use the display to view test results, the 20 most recent results are
available. If you use ProLink III to view results, you can view results for all tests that are in the PC database.
View SMV test results using the display
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view results of previous tests for this meter:
a) Choose Menu > Service Tools > Verification & Calibration > Smart Meter Verification > Read SMV
History.
Pass/Fail results of all tests in the transmitter's SMV database are displayed.
b) To view detailed data for an individual test, select it from the list.
View SMV test results using ProLink III
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view results of previous tests for this meter:
a) Choose Device Tools > Diagnostics > Meter Verification > Run Test.
b) In the SMV Test Definition window, click View Previous Test Results.
c) Click Next.
ProLink III displays a report containing details of the most recent test. The report is automatically
saved to the SMV database. You can print or export the report.
d) To view details of previous tests, click View previous test report.
View SMV test results using the Field Communicator
In addition to test results, the Field Communicator provides a trend chart.
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view detailed results for the current test, choose Service Tools >
Maintenance > Routine Maintenance > SMV > Manual Verification > Most Recent Test Results.
For licensed versions of SMV only, to view results of previous tests:
Measurement support Configuration and Use Manual