Important
Once Main Menu Available has been disabled, you cannot enable it from the local display. Use another
configuration tool, such as ProLink III, to re-enable main menu access from the local display.
7.1.7 Configure totalizer and inventory control
Display Menu > Configuration > Security > Display Security > Totalizers & Inventories
ProLink III Device Tools > Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Behavior
You can enable or disable the operator's ability to start, stop, or reset totalizers or inventories. The totalizer/
inventory control context menus password can also be configured. The context menu is accessed by selecting
the options menu from the process variable screen when a total or inventory is displayed.
These parameters do not affect the operator's ability to start, stop, or reset totalizers or inventories using
another tool.
Procedure
1. Enable or disable Reset Totalizers, as desired.
Default = Enable
2. Enable or disable Start/Stop Totalizers, as desired.
Default = Enable
3. Enable or disable Reset Inventory, as desired.
Default = Disable
4. Enable or disable Start/Stop Inventory, as desired.
Default = Disable
5. If required, configure the totalizer/inventory control context menu password.
Option
Description
No password
(default)
Start, stop or reset totalizer/inventory actions via the context menu do not
require a password.
Password Required When a user selects Start/Stop or Reset from the context menu and the
password is enabled for totalizer/inventory control, the user is required to enter
a password before the action occurs.
If the context menu password option has been set to Password Required, you will be prompted to
enter the password.
Default: AAAA
Range: Any four alphanumeric characters
The display Password Required must be set to At Write or Enter Menu to enable the totalizer/
inventory control context menu password option.