3. Ensure that the valve and channel are powered.
17.3.4 Batch totals are inaccurate
Cause
• Batch target is incorrect
• AOC is inaccurate
• Flow measurement needs adjustment
Recommended actions
1. Verify that the correct batch preset is active.
2. Verify that the batch target is set correctly.
3. If you have recently changed the batch target, run a few batches and wait for the AOC value to adjust.
4. Repeat AOC calibration.
5. Adjust the fixed AOC value.
6. Compare the batch total to a scale reading.
17.3.5 ts BatchRepeatNotSatis
Cause
• AOC is inaccurate
• Inconsistent or leaking valve
• Process is unstable
Recommended actions
1. Inspect the valve and replace if necessary.
2. Repeat AOC calibration.
3. Ensure that line pressure is stable.
4. Ensure that bowl level is approximately constant.
5. Ensure that the pump is not causing unstable flow.
17.4 Concentration measurement problems
17.4.1 Significantly incorrect CM after loading matrix
Cause
The wrong temperature or density unit was configured when the matrix was loaded.
Recommended actions
Set the temperature and density units to the units used when the matrix was built, then reload the matrix.
For custom matrices, contact customer support.
Configuration and Use Manual Troubleshooting
MMI-20025166 March 2019
Configuration and Use Manual 247

Table of Contents

Other manuals for Emerson Micro Motion 5700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion 5700 and is the answer not in the manual?

Emerson Micro Motion 5700 Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion 5700
CategoryTransmitter
LanguageEnglish

Related product manuals