6.7.1 Options for Process Variable Fault Action
Label Description
Display ProLink III Field Communicator
Upscale Upscale Upscale Process variable values indicate that the
value is greater than the upper sensor limit.
Totalizers stop incrementing.
Downscale Downscale Downscale Process variable values indicate that the
value is lower than the lower sensor limit.
Totalizers stop incrementing.
Zero Zero IntZero-All 0 Flow rate variables go to the value that
represents a flow rate of 0 (zero).
Density is reported as0.
Temperature is reported as 0°C , or the
equivalent if other units are used (e.g.,
32°F .
Drive gain is reported as measured.
Totalizers stop incrementing.
Not-a-Number (NAN) Not a Number Not-a-Number Process variables are reported as IEEE
NAN.
Drive gain is reported as measured.
Modbus scaled integers are reported as
Max Int.
Totalizers stop incrementing.
Flow to Zero Flow to Zero IntZero-Flow 0 Flow rates are reported as 0.
Other process variables are reported as
measured.
Totalizers stop incrementing.
None (default) None None (default) All process variables are reported as
measured.
Totalizers increment if they are running.
Configuration and Use Manual Configure advanced options for process measurement
MMI-20025166 March 2019
Configuration and Use Manual 115