Table 9-3: Bit structure of double-precision bytes
Byte Bits Definition
1 SEEEEEEE S=Sign
E=Exponent
2 EEEEMMMM E=Exponent
M=Mantissa
3–8 MMMMMMMM M=Mantissa
5. Optional: Set Additional Communications Response Delay in delay units.
A delay unit is 2/3 of the time required to transmit one character, as calculated for the port currently in
use and the character transmission parameters.
Additional Communications Response Delay is used to synchronize Modbus communications with
hosts that operate at a slower speed than the device. The value specified here will be added to each
response the device sends to the host.
• Default: 0
• Range: 0 to 255
Tip
Do not set Additional Communications Response Delayunless required by your Modbus host.
Configuration and Use Manual Configure digital communications
MMI-20025166 March 2019
Configuration and Use Manual 173