Unit description
Label
Display ProLink III Field Communicator
Grams per minute gram/min g/min g/min
Grams per hour gram/h g/hr g/h
Kilograms per second kg/s kg/sec kg/s
Kilograms per minute kg/min kg/min kg/min
Kilograms per hour kg/h kg/hr kg/h
Kilograms per day kg/d kg/day kg/d
Metric tons per minute MetTon/min mTon/min MetTon/min
Metric tons per hour MetTon/h mTon/hr MetTon/h
Metric tons per day MetTon/d mTon/day MetTon/d
Pounds per second lb/s lbs/sec lb/s
Pounds per minute lb/min lbs/min lb/min
Pounds per hour lb/h lbs/hr lb/h
Pounds per day lb/d lbs/day lb/d
Short tons (2000 pounds) per minute STon/min sTon/min STon/min
Short tons (2000 pounds) per hour STon/h sTon/hr STon/h
Short tons (2000 pounds) per day STon/d sTon/day STon/d
Long tons (2240 pounds) per hour LTon/h lTon/hr LTon/h
Long tons (2240 pounds) per day LTon/d lTon/day LTon/d
Special unit SPECIAL Special Special
Define a special measurement unit for mass flow
Display
Menu > Configuration > Process Measurement > Flow Variables > Mass Flow Settings > Units > SPECIAL
ProLink III Device Tools > Configuration > Process Measurement > Flow > Mass Flow Rate Unit > Special
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > Special Units > Mass Special Units
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer
data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is
calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
Configuration and Use Manual Configure process measurement
MMI-20025166 March 2019
Configuration and Use Manual 33