Options for Discrete Output Source
Option Label State DO voltage
Display PLIII FC Web browser
Enhanced
Event 1–5
(1)
Basic Event x Enhanced
Event x
Discrete
Event x
Enhanced
Event x
ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Event 1–2
(2)
Enhanced
Event x
Event x Event x N/A ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Flow Rate
Switch
Flow Rate
Switch
Flow Switch
Indicator
Flow Switch Flow Rate
Switch
ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Forward/
Reverse
Indicator
Flow
Direction
Forward
Reverse
Indicator
Forward/
Reverse
Flow Direction Forwa
rd
flow
0 V
Rever
se
flow
• Internally powered:
24 VDC
• Externally powered: Site-
specific
Calibration in
Progress
Zero in
Progress
Calibration in
Progress
Cal in
Progress
Zero in
Progress
ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Fault Fault Fault
Indication
Fault
Condition
Present
Fault ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Meter
Verification
Failure
Meter
Verification
Fail
Meter
Verification
Failure
SMV Fail Meter
Verification Fail
ON • Internally powered:
24 VDC
• Externally powered: Site-
specific
OFF 0 V
Configuration and Use Manual Integrate the meter with the control system
MMI-20025166 March 2019
Configuration and Use Manual 159