SMV log
Device UID: 577937183
Device Tag: SUPPLY
Time Zone: GMT -7.00
Date Time LPO Stiff RPO Stiff LPO Mass RPO Mass Damping Drv mA …
8/13/2018 19:27 0.285876 0.289738 0.155294 0.158114 4.41E-05 1.301 …
8/14/2018 7:27 -0.06137 -0.05808 0.154748 0.157556 4.02E-05 1.304 …
8/14/2018 19:27 0.204754 0.20932 0.155185 0.158004 4.35E-05 1.308 …
8/15/2018 7:27 -0.15382 -0.15216 0.154612 0.157416 3.93E-05 1.307 …
8/18/2018 16:27 0.251067 0.251782 0.155217 0.158031 4.34E-05 1.308 …
8/19/2018 19:27 -0.13654 -0.14112 0.154602 0.157396 3.89E-05 1.287 …
8/20/2018 16:27 -0.20837 -0.20671 0.154502 0.157304 3.85E-05 1.291 …
8/21/2018 17:10 -0.11062 -0.11566 0.154641 0.157435 3.84E-05 1.288 …
8/22/2018 10:40 -0.15852 -0.16036 0.154512 0.157308 3.86E-05 1.284 …
8/25/2018 15:40 -0.00172 0.002301 0.154788 0.157599 4E-05 1.295 …
8/27/2018 23:16 0.132787 0.13684 0.155034 0.15785 4.08E-05 1.275 …
8/28/2018 11:16 0.04456 0.046158 0.154845 0.157653 3.99E-05 1.277 …
…
Note
The SMV log displays only in English.
16.1.3 Totalizer history and log
You can configure the transmitter to save totalizer and inventory values at a user-specified interval. You can
then generate a totalizer log. The totalizer log is a ASCII file.
Contents of totalizer log
The totalizer log contains one record for each logged totalizer or inventory value. Each record contains the
following information:
• Default totalizer or inventory name (user-specified names are not used)
• Value and measurement unit
• Timestamp
— Format: Military time
— Time and time zone: Transmitter clock
The totalizer log also contains a line item for each totalizer or inventory reset.
Totalizer history and power-cycles
If the transmitter is rebooted or power-cycled, totalizer history is not affected.
Configuration and Use Manual Log files, history files, and service files
MMI-20025166 March 2019