Table 14-1: SMV abort codes (continued)
Code Description Recommended actions
12 Transmitter in fault and SMV
cannot run
View the Alerts present on the device and take any necessary
actions to clear them before proceeding.
13 No factory air verification Perform factory calibration on air.
14 No factory water verification Perform a factory calibration on water.
15 Drive frequency drift from carrier
frequency
Ensure temperature, flow, and density are stable. Start again.
Other General abort Repeat the test. If the test aborts again, contact customer
service.
14.1.3 Set up SMV automatic execution
You can set up and run a single test at a user-defined future time. You can also set up and run tests
automatically on a regular schedule.
Automatic execution of SMV is managed from the transmitter. You do not need a connection from an
external configuration tool.
Tip
The time between test runs must be between 1 and 1000 hours. The time to the first test run can be any
positive floating number.
Important
Only the 20 most recent SMV results are stored. For SMV Basic versions, the last automatic execution result
can be viewed, while all 20 results can be viewed with SMV licensed. To view or chart these results using an
external tool, you must upload them from the transmitter.
Set up SMV automatic execution using the display
Procedure
1. Choose Menu > Service Tools > Verification & Calibration > Smart Meter Verification > Schedule
SMV.
2. To schedule a single test:
a) Set Hours to 1st Run to the number of hours to elapse before the test is run.
b) Set Hours Between to 0.
3. To schedule a recurring execution:
a) Set Specify Time Until Next Run to the number of days, hours, and minutes to elapse before the
first test is run.
b) Set Specify Time Between Recurring Runs to the number of days, hours, and minutes to elapse
between runs.
4. To disable scheduled execution:
a) Set Specify Time Until Next Run to 0 days, 0 hours, and 0 minutes.
Measurement support Configuration and Use Manual
March 2019 MMI-20025166
204 Micro Motion Model 5700 Transmitters with Configurable Outputs