Emerson Micro Motion 5700 - Configure Batch Resets; Configure Batch Presets Using the Display

Emerson Micro Motion 5700
356 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Two-stage batching is a batch that is run on a gas and liquid mixture flowing through a pipeline.
Parameter Value
Number of Stages Set to 2.
Setting this parameter to 1 disables two-stage batching.
Configure Presets by
% Target The values for Open Primary, Open Secondary, Close
Primary, and End Warning are each configured as a
percent of target.
Quantity The values for Open Primary and Open Secondary are
each configured as a quantity at which the valve should
open. The values for Close Primary and End Warning
are each configured as a quantity that is subtracted
from the target.
5.3.2 Configure batch resets
Configure batch presets using the display
A batch preset is a predefined group of batching parameters. You can define and save up to six batch presets.
When you run a batch, you must choose a preset. You can modify the batch target before you run the batch.
All other settings are fixed.
Procedure
1. Choose Menu > Configuration > Process Measurement > Batcher > Presets.
2. Select the preset that you want to configure.
3. Set Preset Status as desired.
Option
Description
Enabled (On) The preset is available for use. You can select it to run a batch.
Disabled (Off) The preset is not available for use. You cannot select it to run a batch. You can always
enable it at a later time.
Preset 1 is always enabled and cannot be disabled.
4. Set Name to the name to be used for this preset.
The name can be up to eight characters in length. Valid characters include A–Z and 0–9.
5. Set Target to the size of the batch.
When the transmitter has measured the specified amount and applied the selected form of AOC, it
closes the valve.
6. Optional: Set End Warning to the point in the batch at which a Modbus coil will be set.
Configuration and Use Manual Configure process measurement applications
MMI-20025166 March 2019