Option Flow direction Totalizer and inventory behavior
Reverse Totals do not change
Reverse Only Forward Totals do not change
Reverse Totals increment
Bidirectional Forward Totals increment
Reverse Totals decrement
Absolute Value Forward Totals increment
Reverse Totals increment
Important
Actual flow direction interacts with Sensor Flow Direction Arrow to determine the flow direction that
the transmitter uses in processing. See the following table.
Table 6-1: Interaction between actual flow direction and Sensor Flow Direction Arrow
Actual flow direction Setting of Sensor Flow Direction
Arrow
Flow direction sent to outputs
and totalizers
Forward (same direction as Flow
arrow on sensor)
With Arrow Forward
Against Arrow Reverse
Reverse (opposite from Flow arrow
on sensor)
With Arrow Reverse
Against Arrow Forward
4. Optional: Set User Name to the name you want to use for the inventory or totalizer.
User Name can have a maximum of 16 characters.
The transmitter automatically generates a name for each totalizer and inventory, based on its source,
direction, and type.
Example
Totalizer Source=Mass Flow
Totalizer Direction=Forward Only
Totalizer name=Mass Fwd Total
Example
Inventory Source=Gas Standard Volume Flow
Inventory Direction=Bidirectional
Inventory name=GSV Bidir Inv
The specified name is used on the transmitter display and on all interfaces that support it. If User Name
contains only spaces, the transmitter-generated name is used. Not all interfaces support totalizer and
inventory names.
Configure advanced options for process measurement Configuration and Use Manual
March 2019 MMI-20025166
112 Micro Motion Model 5700 Transmitters with Configurable Outputs