12.4.3 Start, stop, and reset totalizers using the Field
Communicator
You can start and stop each totalizer or inventory independently. You can start and stop all totalizers and
inventories as a group. You can reset each totalizer or inventory independently. You can reset all totalizers as
a group. You can reset all inventories as a group.
When a totalizer or inventory is started, its value increases or decreases depending on the interaction of the
flow direction parameters. It continues tracking flow until it is stopped.
When a totalizer or inventory is reset, its value is set to 0. You can reset a totalizer or inventory while it is
started or while it is stopped.
Procedure
• To start or stop a single totalizer:
a) Choose Overview > Totalizer Control > Totalizers 1-7.
b) Select the totalizer that you want to start or stop.
c) Choose Start or Stop.
• To start or stop a single inventory:
a) Choose Overview > Totalizer Control > Inventories 1-7.
b) Select the inventory that you want to start or stop.
c) Choose Start or Stop.
• To start or stop all totalizers and inventories as a group:
a) Choose Overview > Totalizer Control.
b) Click Start Totalizers or Stop Totalizers.
• To reset a single totalizer:
a) Choose Overview > Totalizer Control > Totalizers 1-7.
b) Select the totalizer that you want to reset.
c) Choose Reset.
• To reset a single totalizer:
a) Choose Overview > Totalizer Control > Inventories 1-7.
b) Select the inventory that you want to reset.
c) Choose Reset.
• To reset all totalizers as a group, choose Overview > Totalizer Control > Reset All Totals.
• To reset all inventories as a group, choose Overview > Totalizer Control > Reset All Inventories.
Transmitter operation Configuration and Use Manual
March 2019 MMI-20025166
194 Micro Motion Model 5700 Transmitters with Configurable Outputs