Emerson Micro Motion 5700 - Page 196

Emerson Micro Motion 5700
356 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Optional: Pause the batch at any time.
Using the display: Choose Menu > Batcher > Run Batch > Batch Control and select Pause.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher and click Pause
Batcher.
Pausing a batch allows you to resume it later.
6. Optional: Resume the batch after it has been paused.
Using the display: Choose Menu > Batcher > Run Batch > Batch Control and select Resume.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher and click Resume
Batch.
7. Optional: End the batch manually at any time before it reaches the target.
Using the display: Choose Menu > Batcher > Run Batch > Batch Control and select End.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher and click End Batch.
If you end a batch manually, you will not be able to resume it. You can end the batch while it is running
and while it is paused.
8. If you did not end the batch manually, wait until the batch ends automatically.
Using the display: Choose Menu > Batcher > Run Batch > Batch Control.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher.
The batch ends automatically when it reaches the target or when it reaches the configured maximum
fill time.
9. Optional: Inhibit batch.
This function is not available from the display.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher.
While the Inhibit Batch input is active, a batch will not be allowed to start. Attempting to start a batch
while the batch is inhibited will cause the Start Not Okay alarm to be posted.
10. Optional: Inhibit totalizer.
This function is not available from the display.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher.
When the Inhibit Totalizer input is active, the Batch Total will not increment. A timeout alarm will
occur if the Batch Total is inhibited beyond the max batch time. The batch inventory for the current
preset will not increment while Inhibit Totalizer is active.
11. Optional: Inhibit flow.
This function is not available from the display.
Using ProLink III: Choose Device Tools > Application Control > Run Batcher.
When the Inhibit Flow input is active, the batch will ignore the flow rate of the assigned flow source,
allowing a batch to end. The batch total will not increment when flow is inhibited.
Operation using the batcher Configuration and Use Manual
March 2019 MMI-20025166
196 Micro Motion Model 5700 Transmitters with Configurable Outputs