g) Start the simulation.
h) Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct
the discrepancy by trimming the output.
i) Choose Exit.
2. Test the Frequency Output(s).
a) Choose Menu > Service Tools > Output Simulation and select the frequency output to test.
Note
If you have configured both Channel B and Channel D as Frequency Output 2, be sure to test
both loops.
b) Set Simulation Value to 1.
c) Start the simulation.
d) Read the frequency signal at the receiving device and compare it to the transmitter output.
e) Choose New Value.
f) Set Simulation Value to 14500.
g) Start the simulation.
h) Read the frequency signal at the receiving device and compare it to the transmitter output.
i) Choose Exit.
3. Test the Discrete Output(s).
a) Choose Menu > Service Tools > Output Simulation and select the discrete output to test.
b) Set Simulation Value to ON.
c) Start the simulation.
d) Verify the signal at the receiving device.
e) Choose New Value.
f) Set Simulation Value to OFF.
g) Start the simulation.
h) Verify the signal at the receiving device.
i) Choose Exit.
4. Test the Discrete Input.
a) Set the remote input device to ON.
b) Choose Menu > Service Tools > Service Data > View I/O Levels and read the state of the discrete
input.
c) Set the remote input device to OFF.
Troubleshooting Configuration and Use Manual
March 2019 MMI-20025166
292 Micro Motion Model 5700 Transmitters with Configurable Outputs