Table 17-7: Velocity measurement problems and recommended actions (continued)
Problem Possible causes Recommended actions
Inaccurate velocity
reading
Wiring problem
Inappropriate measurement unit
Incorrect flow calibration factor
Incorrect density calibration factors
Incorrect grounding
Two-phase flow
Problem with receiving device
Incorrect sensor zero
Verify that the measurement units are
configured correctly for your application.
Zero the meter.
Check the grounding of all components.
Check for two-phase flow.
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
Replace the core processor or transmitter.