EasyManua.ls Logo

Flexim FLUXUS G704 - Transducer Distance

Flexim FLUXUS G704
257 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
104 UMG70XV3-4EN 12.01.2009
11 Displaying the Measured Values
11.5 Transducer Distance
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses in the
upper line, followed by the entered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during measurement (e.g. due to tem-
perature fluctuations).
A deviation from the optimum transducer distance (here: -0.4 mm) will be compensated
internally.
R flow profile
information about the flow profile based on the Reynold's number
T fully turbulent flow profile
L fully laminar flow profile
the flow is in the transition range between laminar and turbulent flow
? unknown, can not be calculated
F flow velocity
comparison of the measured flow velocity with the flow limits of the system
ok, flow velocity is not in a critical range
the flow velocity is higher than the actual limit
the flow velocity is lower than the actual cut-off flow (even if it is not set to zero)
0 the flow velocity is in the offset range of the measuring method
? unknown, can not be measured
By pressing key , it is possible to scroll to the display of
the transducer distance during measurement.
Note! Never change the transducer distance during measurement!

L=(51.2) 50.8 mm
54.5 m3/h
DISP

Table of Contents

Related product manuals