136 UMG70XV3-4EN 12.01.2009
14 Libraries
14.3.5 Input of Steam and Concentration Coefficients
Use the program FluxKoef (option).
14.4 Deleting a User Defined Material/Medium
To delete a user defined material/medium, proceed as follows:
Select in the program branch SPECIAL FUNCTION the menu item INSTALL MATERIAL
or INSTALL MEDIUM. Press ENTER.
If the extended library is activated, press ENTER until the request for deleting is dis-
played.
14.5 Arrangement of the Material/Medium Scroll List
The materials and media to be displayed in the program branch PARAMETER will be ar-
ranged in the material scroll list and in the medium scroll list.
Enter the 10 heat flow coefficients: a0...a4, r0...r4. Press
ENTER after each input.
Select YES to store the heat flow coefficients. Press EN-
TER.
Note! The entered coefficients will not be checked. Absurd values can re-
sult in wrong measured values or in permanent system errors.
Select DELETE. Press ENTER.
Select the material/medium to be deleted. Press ENTER.
Select YES or NO. Press ENTER.
Note! User defined materials/media will always be displayed in the scroll
lists of the program branch PARAMETER.
Select SPECIAL FUNCTION\SYSTEM SETTINGS\LI-
BRARIES. Press ENTER.
Heat-flow Coeffs
Save? no >YES<
INSTALL MATERIAL
edit >DELETE<
USER MATERIAL
#01: Polystyrol
SYSTEM settings
Libraries