UMG70XV3-4EN 12.01.2009 147
16 SuperUser-Mode
16.3 Malfunctions in SuperUser Mode
As the SuperUser mode operates without any plausibility check, absurd entries may re-
sult in an automatic switching-off of the flowmeter or in a crash of the internal software.
An absurd input is e.g. 0 (zero) for the number of sound paths or 0.1 mm for the outer
pipe diameter.
Switch on the flowmeter again and reactivate the SuperUser mode. Use RESET, if nec-
essary by pressing keys BRK, C and ENTER simultaneously.