6 Installation of FLUXUS G704
6.7 Connection of the Outputs
• Prepare the output cable. Connect the leads with the terminals of the outputs as dis-
played (see Fig. 6.12 and Table 6.11).
Fig. 6.12: Connection of the outputs
• Close the flowmeter: Screw the cover on the housing.
Configure the outputs (see chapter 18). The terminals
to be used for the connection of the output are dis-
played at the end of the configuration dialog (here: P1+
and P1- for the active current loop).
Table 6.11: Circuits of the outputs
output FLUXUS terminal circuits
active
current loop
Px+
(red)
(black)
Px-
R
LOAD
< 500
semi-active
current loop
operated as
active current
loop
Px+
(red)
(black)
Px-
R
LOAD
< 50
I1 active loop
Terminal:P1+,P1-