EasyManuals Logo

Flexim FLUXUS G704 User Manual

Flexim FLUXUS G704
257 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
52 UMG70XV3-4EN 12.01.2009
6 Installation of FLUXUS G704
6.7 Connection of the Outputs
Prepare the output cable. Connect the leads with the terminals of the outputs as dis-
played (see Fig. 6.12 and Table 6.11).
Fig. 6.12: Connection of the outputs
Close the flowmeter: Screw the cover on the housing.
Configure the outputs (see chapter 18). The terminals
to be used for the connection of the output are dis-
played at the end of the configuration dialog (here: P1+
and P1- for the active current loop).
Table 6.11: Circuits of the outputs
output FLUXUS terminal circuits
active
current loop
Px+
(red)
(black)
Px-
R
LOAD
< 500
semi-active
current loop
operated as
active current
loop
Px+
(red)
(black)
Px-
R
LOAD
< 50
I1 active loop
Terminal:P1+,P1-
outputs
(KL4)
+
-
mA
+-
+
-
mA
+-

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS G704 and is the answer not in the manual?

Flexim FLUXUS G704 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS G704
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals