146 UMG70XV3-4EN 12.01.2009
16 SuperUser-Mode
16 SuperUser-Mode
The SuperUser mode allows experimental work. Features of the SuperUser mode are:
• Defaults will not be observed.
• There are no plausibility checks when parameters are being entered.
• There is no check whether the entered parameters are within the limit determined by
physical laws and technical data.
• The cut-off flow is not active.
• A value for the number of sound paths has to entered.
It is possible to change the lower limit of the inner pipe diameter for a given transducer
type without the activation of the Super User mode.
16.1 Activating/Deactivating
Press key C. Enter HotCode 071049 .
Press key C. Enter HotCode 071049 again to deactivate the SuperUser mode.
The SuperUser mode will be deactivated by switching off the flowmeter, too.
16.2 Transducer Parameters
In SuperUser mode, the menu item TRANSDUCER TYPE will be displayed at the end of
parameter input even if the transducers are detected by the flowmeter.
It is displayed that the SuperUser mode is activated. Press
ENTER. The main menu will be displayed.
It is displayed that the SuperUser mode is deactivated.
Press ENTER. The main menu will be displayed.
Press ENTER.
or
Select SPECIAL VERSION to edit the transducer parame-
ters. Press ENTER.
If SPECIAL VERSION is selected, the transducer parame-
ters have to be entered.
The transducer parameters have to be provided by the
transducer manufacturer. Press ENTER after each input.
SUPERUSER MODE
*IS ACTIVE NOW*
SUPERUSER MODE
IS PASSIVE NOW
Transducer Type
Special Version