EasyManua.ls Logo

Flexim FLUXUS G704 - Page 183

Flexim FLUXUS G704
257 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UMG70XV3-4EN 12.01.2009 183
A Technical Data
Shear Wave Transducers (for ATEX zone 2, FM or without Explosion Protection)
technical type GDK1N52 GDM1N52 GDP1N52 GDG1N52
order code GSK-NA2TS
GSK-NF2TS
GSK-NNNTS
GSM-NA2TS
GSM-NF2TS
GSM-NNNTS
GSP-NA2TS
GSP-NF2TS
GSP-NNNTS
GSG-NA2TS
GSG-NF2TS
GSG-NNNTS
transducer frequency MHz 0.5 1 2 0.2
medium pressure
1
min. extended bar 20 20 20 20
min. bar metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
outer pipe diameter
2
min. extended mm 70 30 15 250
min. recommended mm 80 40 20 380
max. recommended mm 500 80 40 810
max. extended mm 720 120 60 1100
pipe wall thickness
min. mm 5 2.5 1.5 14
max. mm - - - -
material
housing PEEK with
stainless steel
cap304
(1.4301)
stainless steel
304 (1.4301)
stainless steel
304 (1.4301)
PEEK with
stainless steel
cap304
(1.4301)
contact surface PEEK PEEK PEEK PEEK
degree of protection
according to EN 60529
IP 67 IP 67 IP 67 IP 67
transducer cable
type 1699 1699 1699 1699
length m 5 4 4 5
dimensions
length l mm 126.5 60 60 129.5
width b mm47303047
height h mm 55.9 33.5 33.5 66.4
dimensional drawing
operating temperature
min. °C -40 -40 -40 -40
max. °C +130 +130 +130 +130
continued on next page
h
b
l
h
b
l
h
b
l
h
b
l

Table of Contents