UMG70XV3-4EN 12.01.2009 195
A Technical Data
Lamb Wave Transducers (for ATEX Zone 2, FM or without Explosion Protection)
technical type GRH1N52 GRK1N52 GRG1N52
order code GLH-NA2TS
GLH-NF2TS
GLH-NNNTS
GLK-NA2TS
GLK-NF2TS
GLK-NNNTS
GLG-NA2TS
GLG-NF2TS
GLG-NNNTS
transducer frequency MHz 0.3 0.5 0.2
medium pressure
1
min. extended bar metal pipe: 10 metal pipe:
10 (> DN 120),
5 (< DN 120)
metal pipe: 10
min. bar metal pipe: 15
plastic pipe: 1
metal pipe:
15 (> DN 120),
10 (< DN 120)
plastic pipe: 1
metal pipe: 15
plastic pipe: 1
outer pipe diameter
2
min. extended mm 120 60 190
min. recommended mm 140 80 220
max. recommended mm 600 300 900
max. extended mm 1000 500 1600
pipe wall thickness
min. mm 7 4 11
max. mm 15 9 23
material
housing PPSU with
stainless steel
cap304
(1.4301)
PPSU with
stainless steel
cap304
(1.4301)
PPSU with
stainless steel
cap304
(1.4301)
contact surface PPSU PPSU PPSU
degree of protection
according to EN 60529
IP 67 IP 67 IP 67
transducer cable
type 1699 1699 1699
length m 5 5 5
dimensions
length l mm 128.5 128.5 128.5
width b mm474747
height h mm 69.9 69.9 69.9
dimensional drawing
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
continued on next page