FT-111, User Manual Rev. 2.0.0, June 2022 Page 136 of 137
- Press enter key and re-enter the footers.
- Change E
2
ROM (U6).
- Press enter key and re-enter the fieldbus
setup.
- Change E
2
ROM (U6).
Set point (limit values )
checksum error.
- Press enter key and re-enter the set point
(limit) values.
- Change E
2
ROM (U6).
- Change CR2032 battery.
- Change the main board.
Checksum error of EPL
printout format.
- Press enter key and re-load the EPL code
- Change E
2
ROM (U6).
ADC output is not changed for
the span calibration.
- Recalibrate the scale.
- Change mainboard.
The load cell output is
decreased after loading.
- Check load cell connections.
- Check test weight loading.
- Recalibrate the scale.
- Change mainboard.
The load is not enough for
span calibration.
The load is not stable at
calibration.
- Wait until scale become stable.
- Check grounding wiring.
- Recalibrate the scale.
- Change mainboard.
Alibi memory SD card is not
installed.
- Disable Alibi memory if not required.
- Check Alibi SD card.
- Change mainboard.
Alibi SD card is not supplied
from FLINTEC.
- Install Flintec Alibi SD card.
Alibi memory serial number
error. The new alibi SD card is
installed.
- Format the alibi memory SD card.
Refer to parameter 816.
Alibi memory could not be
initialized.
- Check alibi memory SD card
- Change main board.
- Check alibi memory records.
- Change Alibi memory SD card.
- Change main board.
Alibi information CSUM error.
Wrong SD card at alibi
memory.
- Check data format of other interfaces.
- Other interfaces should not be Modbus.
Re-verification of the scale.
- Reverify the scale after checking the
scale hardware, load cells, performance
etc.