FT-111, User Manual Rev. 2.0.0, June 2022 Page 78 of 137
11.2.2 Activating Alibi Memory
1. Go to the parameter [ 8-- METROLOGY] main block in set up.
2. Press key 2 times to access parameter 811.
3. Press key after selecting YES in the parameter.
4. Press key until [ SAVE : YES ] prompt seen on the display and press key to save the
changes into the memory. [ E42 NEW ALIBI ] message appears.
5. Re-enter the set-up and format the SD memory card in parameter 816.
6. Press key until [ SAVE : YES ] prompt seen on the display and press key to exit from
set up.
11.2.3 How is the Data Recorded into the Alibi Memory?
After the following actions the transferred data is recorded into the alibi memory:
1. Press key to print the label,
2. Automatic printing (automatic print data transferring),
3. Interface command ASCII < P >,
4. BSI commands which request the stable weight,
5. Print commands of fieldbus interfaces,
11.2.4 Recall Data from Alibi Memory
RELATED PARAMETERS: Main block 8--.
You can fulfil your weight data recording obligations with the alibi memory in certified operation without
having paper archive. The weight data is saved into the alibi memory after every weighing operation
together with date and time. The alibi record number may also be found on the printout data.
After the following actions the data is recorded into the alibi memory:
1. Press key to print the label,
2. Automatic printing (automatic data transferring),
3. Interface command ASCII < P >,
4. BSI commands which request the stable weight,
5. Print commands of fieldbus interfaces,
6. Digital < Print > input.
The alibi memory recording principle is a loop memory which deletes the oldest record after the capacity
limit 99 999 has been reached to overwrite the latest weighing. You can quickly access to the record of
specific weighing by entering the suitable search criteria.
65.
Then select one of the suitable criteria:
Alibi record number, date, net weight, gross weight or tare weight. You can also transfer all alibi memory
records to your PC.
Access to Alibi records:
1. Go to the parameter [ 8-- METROLOGY] main block in set up.
2. Press key 3 times and key to access parameter 813.
3. Select the suitable criteria by pressing keys which are NUM (Alibi number), DATE (weighing
date) or CN (consecutive number).